Wavefront technology in eye exams
Wavefront technology developed for custom LASIK may soon be used routinely by eye doctors to better diagnose vision problems in eye exams, perhaps making the familiar eye chart obsolete.
Most people have had eye exams with a device called a phoropter, which contains many lenses of different powers. An ophthalmologist or optometrist changes the lenses in front of your eyes, asking which lens produces the best image.
With this conventional approach, information you give the eye doctor is very subjective, based more on what you think you see instead of what you actually see. But a wavefront measurement is objective, because vision errors can be identified automatically by the way light waves travel through the eye.
Someday, these detailed wavefront measurements may replace conventional eyeglass or contact lens prescriptions, which describe vision problems only in terms of the eye's nearsightedness, farsightedness and astigmatism.
Just as custom (or "wavefront-guided") LASIK has the potential for producing sharper vision than conventional LASIK, glasses and contact lenses made with this advanced technology may also produce better visual clarity than their conventional counterparts.
For more about the differences between conventional and wavefront-guided eyewear prescriptions, read: How high-definition eyeglass lenses offer sharper vision.
WHEN WAS YOUR LAST EYE EXAM? Find an eye doctor near you (and ask about wavefront eye exams.
What is a wavefront eye exam?
So exactly what does "wavefront" mean?
Imagine light traveling as a bundle of rays. If you draw lines perpendicular to the tips of a bundle of light rays, you get what is called a wavefront map. In an eye with perfect vision, the wavefront is perfectly flat, but the wavefront of an imperfect eye is irregular.
The types of distortions this wavefront acquires as it travels through the eye provide valuable information about vision errors and how to correct them.
What is wavefront technology (aberrometry)?
Aberrometry measures the way a wavefront of light passes through the cornea and the crystalline lens, which are the refractive (light focusing) components of the eye. Distortions that occur as light travels through the eye are called aberrations, representing specific vision errors.
Wavefront technology, or aberrometry, diagnoses both lower- and higher-order vision errors represented by the way the eye refracts or focuses light.
Higher-order aberrations are more complex vision errors, whereas lower-order aberrations are more common vision errors such as nearsightedness, farsightedness and astigmatism.
Wavefront and higher-order aberrations
Previously, with conventional methods of eye examinations, only lower-order vision errors could be diagnosed and treated. Higher-order aberrations such as coma, trefoil and spherical aberration were largely ignored by eye care professionals because their impact on vision was believed at the time to be slight and because no feasible means existed to precisely identify or correct them.
Now that higher-order aberrations can be accurately defined by wavefront technology and corrected by new kinds of spectacles, contact lenses, intraocular lenses and refractive surgery, they have become more important factors in eye exams.
Higher-order aberrations started receiving more attention because they were identified as sources of visual side effects after LASIK and other types of refractive surgery — causing halos, ghost images and other visual symptoms.
Newer wavefront-guided lasers used in vision correction surgery can reduce certain higher-order aberrations, which may improve visual performance, especially for low-light activities such as driving at night.
How is a wavefront measurement performed?
A wavefront diagnosis of the eye is performed with an aberrometer.
With most aberrometers, you place your chin on a chin rest. Next, you are asked to peer into the device and focus your eyes on a point of light. After a few seconds of silence (no feedback will be required of you), the wavefront measurement is completed and a wavefront map of your eye prints out for your eye care professional.
In general, an aberrometer uses a three-step process:
Because a wavefront passes through the opening in the front of the eye (pupil), your pupil diameter is measured. This measurement is used to derive a reference wavefront shape representing a theoretically perfect eye that has the exact same pupil size as yours.
A beam of light is projected into your eye and is reflected back by the retina. The wavefront of this reflected light is captured by the aberrometer. Because no eye is optically perfect, all such wavefronts will contain at least some distortions.
An aberration map of the eye is created by comparing the shape of the captured wavefront with that of the pre-programmed reference wavefront, measuring all points of difference between the two. Then a wavefront map of your eye, sometimes called an "optical fingerprint" since no two eyes produce the exact same map, is created.
What does your eye's wavefront map mean?
When you start to interpret the results of your eye's wavefront map, keep in mind that the reference shape used for comparison is flat or two-dimensional.
This flat, circular plane represents a theoretically perfect (emmetropic) eye that exactly matches the diameter of your eye's pupil.
Your eye's actual three-dimensional wavefront map is created through comparison with this theoretically perfect, flat wavefront map. Three-dimensional distortions occur in the actual wavefront map based on how individual light rays are altered as they travel through the eye.
Imperfections in the cornea and lens can alter the path of light rays by causing them to speed up, slow down or incorrectly bend (refract).
Other eye problems also can cause higher-order aberrations, including dry eye that alters the eye's tear film, which also helps refract light rays. Even cataracts clouding the eye's natural lens or scarring of the eye's surface from eye surgery, disease or trauma can cause higher-order aberrations and focusing problems.
Based on the many different patterns that can occur in a wavefront, more than 60 different aberrated shapes have been identified and classified as vision errors.
Wavefront measurements
Wavefront measurements that identify vision errors can be calculated in different ways. Currently, the method of choice is a mathematical expression called a Zernike polynomial.
Defined by Dutch physicist Fritz Zernike in 1934, this polynomial is well suited to the task because it is designed to meet the requirements of a circle.
Each Zernike polynomial describes the aberration existing at a specific point on a wavefront of light, after it passes through an eye. The sum of the Zernike polynomials equals a full description of all the aberrations or total refractive error in a given eye.
This finding is delivered in the form of a Zernike prescription and a topographic map, a detailed drawing of the shape of the aberrated wavefront. Together, they make up an aberration map or unique "optical fingerprint" that identifies the eye's vision errors in detail.
How your wavefront map is used
The wavefront map is a complete, accurate description of all aberrations affecting your eye. Your eye doctor uses your wavefront map as a blueprint for custom-designing your vision correction, whether with refractive surgery such as LASIK, intraocular lenses, contact lenses or glasses.
Remember that you can never achieve a perfect wavefront, because no eye is perfect. Even if you are diagnosed with a certain number of higher-order aberrations, this is not necessarily cause for concern, unless the degree of aberration causes significant vision problems such as difficulty seeing at night.
Page published on Wednesday, February 27, 2019